• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT; MITRACLIP DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT; MITRACLIP DELIVERY SYSTEM Back to Search Results
Catalog Number CDS0501
Device Problems Break (1069); Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Date 10/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the inability to connect the stopcock to the flush port and the broken flush port.It was reported that during preparation of the clip delivery system (cds), it was noted that a stopcock could not be connected to the bottom flush port on the delivery catheter (dc) handle.Upon further inspection, the port was noted to be broken.The cds was not used in the anatomy and was replaced.A new cds was used to complete the procedure.There was no clinically significant delay in the procedure and no patient involvement.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from the lot.All available information was investigated and a definitive cause for the reported broken flush port in this incident could not be determined.The reported issue of the stopcock not able to tighten onto the bottom delivery catheter (dc) flush port was due to the broken flush port.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT
Type of Device
MITRACLIP DELIVERY SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8046811
MDR Text Key126513178
Report Number2024168-2018-08557
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2019
Device Catalogue NumberCDS0501
Device Lot Number80515U131
Was Device Available for Evaluation? No
Date Manufacturer Received01/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-