The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.No corrective actions will be taken at this time.A review of the manufacturing records indicated that the product met specifications upon release.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on a monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.There was no patient compromise noted in this event.It is unknown whether user or procedural factors played a role in this event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.The 510k is not known as this is a custom defined product.
|
As reported, during use of this disposable pressure transducer combo kit, variances in invasive blood pressure (ibp) and non invasive blood pressure (nibp) were observed.Systolic differences were as much as 30mmhg.The line was checked and no abnormalities were observed.A square wave test showed a normal trace and no alarm was displayed.A print out of the pressure was carried out to verify the pressure waveform, which was found to be ¿overshooting¿.The patient was not treated due to the inaccurate values nor was patient injury reported.The patient was post coronary artery bypass grafting.Complete patient demographics were not able to be obtained.Unfortunately, the device was not available as the hospital discarded the device after removal.
|