MAUDE Adverse Event Report: DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number UNKNOWN
Device Problems Malposition of Device (2616); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A review of the image in the article confirms the iab migration.This device was not used in accordance with any iab instructions as it was reported to be an axillary insertion.The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Event Description
This complaint was captured via literature (published article).It was mentioned that total duration intra-aortic balloon (iab) pump (iabp) therapy was 27 days.The position of iab catheter was checked using chest x-ray and it showed iabp markers at t5 and t9 levels respectively.Abdominal x-ray was taken and it showed distal iabp marker at t9 level.It was confirmed via fluoroscopy that iabp proximal end was bent and flopped over itself into ascending aorta.The iab was exchanged to continue therapy.There was no reported injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key8058050
MDR Text Key128211796
Report Number2248146-2018-00649
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age YR
Date Manufacturer Received10/19/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age56 YR
-
-