Model Number MCP00703279
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/02/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).A field service technician was sent out for investigation on (b)(6) 2018 and ran pump for almost 1.5 hours with no stops.Technician adjusted the speed, changed the settings (cpl, free, etc.) and the pump kept running with no issues.Belts and pulleys were inspected with no errors found.The failure could not be reproduced.The technician also stated that the console of the hl20 displayed "em07" (failed +5v test) on startup prior to treatment, but the hospital used the console anyway for the surgery.This may have caused the pump stop.Since this pump is not under getinge care, no further information about the console is available.The failure could not be confirmed.According to service order #(b)(4) no errors were found and no parts were replaced.This complaint will be closed based upon the within oxando generated documentation of the service carried out.If further documentation will get available and adds further relevant information this complaint will be re-opened, documentation updated and further necessary actions will be carried out if applicable.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
It was reported that the twin pump of an hl20 stopped during a case.Pump was reset and case was continued.Patient was not injured.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key8059944
MDR Text Key127075085
Report Number8010762-2018-00301
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00703279
Device Catalogue Number701028694
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1