Model Number 3650
Device Problems Smoking (1585); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 10/30/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the programmer started smoking and fizzing in clinic.The programmer was switched off.The programmer still works, but makes a buzzing sound.No patient was involved in this incident.
 
Manufacturer Narrative
The field complaint of a buzzing sound was verified, but there were no indications of burn damage.During analysis, the programmer was plugged in add powered on.Once booted, there was noise coming from the speakers.The programmer was disassembled and inspected for signs of burning or smoke, but none were found.An internal anomaly was found to be the cause of the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8061648
MDR Text Key126911527
Report Number2938836-2018-12132
Device Sequence Number1
Product Code KRG
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-