Model Number 10-88-00 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that the magnet of a s5 mast roller pump cover was missing during setup.There was no patient involvement.
|
|
Manufacturer Narrative
|
Livanova field service representative was dispatched onsite to investigate.The reported issue could be confirmed.The pump cover was replaced with a new one and subsequent functional verification testing found no further issues.The device was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the issue.A corrective action was implemented for that issue.
|
|