• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OTTO BOCK HEALTHCARE LP OTTO BOCK AXTION® PROSTHETIC FOOT; 1E56 AXTION® PROSTHETIC FOOT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OTTO BOCK HEALTHCARE LP OTTO BOCK AXTION® PROSTHETIC FOOT; 1E56 AXTION® PROSTHETIC FOOT Back to Search Results
Model Number 1E56
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/15/2018
Event Type  malfunction  
Manufacturer Narrative
Submission was originally attempted on 11/13/2018.It was discovered on 11/13/2018 that the original submission was not accepted due use of an incorrect icsr version.Icsr r2 is now required.I re-installed esubmitter in hopes the problem is resolved and my submission will be accepted.The cdrh emdr help desk suggested the re-installing of the esubmitter would resolve this issue.(b)(4).
 
Event Description
The end-user was walking into a movie theatre and heard a crack.Immediately the end-user realized the pyramid adapter had cracked causing the foot to become disconnected from their pylon.The end-user avoided falling and was able to get back to his home safely.He used a back up foot until his axtion foot was replaced.No fall or injury resulted.All information reported by the customer and end-user indicates the foot was used as instructed and within proper indications.The failure occurred within the intended use-life for this device.Out of an abundance of caution, this complaint is being reported as an mdr since this confirmed failure has demonstrated the risk of causing serious injury to the end-user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OTTO BOCK AXTION® PROSTHETIC FOOT
Type of Device
1E56 AXTION® PROSTHETIC FOOT
Manufacturer (Section D)
OTTO BOCK HEALTHCARE LP
3820 west great lakes drive
west valley city UT 84120
Manufacturer (Section G)
OTTO BOCK HEALTHCARE LP
3820 west great lakes drive
west valley city UT 84120
Manufacturer Contact
stephen anderson
3820 west great lakes drive
west valley city, UT 84120
8019746676
MDR Report Key8071584
MDR Text Key129188015
Report Number1721652-2018-00003
Device Sequence Number1
Product Code ISH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1E56
Device Catalogue Number1E56=N29-5-P/0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2018
Date Manufacturer Received10/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Weight120
-
-