STERIS CORPORATION - MONTGOMERY 3085SP SURGICAL TABLE Back to Search Results
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/18/2018
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the surgical table.The technician was informed that after the patient procedure, an employee tried to lower the surgical table, so they could transfer the patient from the surgical table to a transport bed.The table would not lower, so the employees lifted the patient from the surgical table to the transport bed.The technician inspected the surgical table and found that the table's column cover was damaged and misaligned.The facility declined repairs from the technician and the surgical table has not been returned to service.The 3085sp surgical table was installed in 1998 making this unit approximately 20 years old and is not under steris service agreement for maintenance activities.The user facility is responsible for all maintenance activities.The 3085sp surgical table operator manual (1-2) states: "safe and reliable operation of this equipment requires regularly scheduled preventative maintenance, in addition to the faithful performance of routine maintenance.Contact steris to scheduled preventative maintenance." additionally, the 3085sp surgical table operator manual (1-2) states: "repairs and adjustments to this equipment must be made only by fully qualified service personnel.Nonroutine maintenance performed by inexperienced, unqualified personnel or installation of unauthorized parts could cause personnel injury, invalidate the warranty, or result in costly damage.Contact steris for service options." no additional issues have been reported.
 
Event Description
The user facility reported that their 3085sp surgical table would not lower to the desired height at the end of a patient procedure.No injury or procedure delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3085SP SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key8075566
MDR Text Key129068829
Report Number1043572-2018-00092
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/18/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-