Results: the cat6 was kinked approximately 1.0 cm and 9.5 cm from the hub.The distal tip was ovalized.Conclusions: evaluation of the returned cat6 revealed a kinked device.If the device is forcefully withdrawn from the packaging at extreme angles or otherwise forcefully handled, damage such as a kink may occur.Further evaluation revealed an ovalization on the distal tip.If the device is forcefully gripped or pinched, damage such as an ovalization may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The product lot number was not provided; therefore, the manufacturing records could not be reviewed.
|
During preparation for a thrombectomy procedure, the hospital staff noticed that an indigo system aspiration catheter 6 (cat6) was kinked upon removal from the package.The damage to the cat6 was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using a new cat6.
|