The device was returned for evaluation.With respect to the returned unit, it has passed all specific functional testing requirements, except for the lock having rotational movement.However, when unit is not under pressure, this would not have caused a slippage.When unit is properly positioned and put under pressure, device would not have slipped.This device exceeds its expected life of 7 years.Complaint is not confirmed.The root cause is unknown.However, review of the foreseeable sequence of events indicated in risk management file indicates the most likely reason for reported failure of "unintended movement of the patient¿s head¿ is: clinician or staff applied an inadequate load to the patient¿s head; clinician used unapproved skull pins; index locking mechanism was not fully engaged; clinician/staff applied a clamping load greater than the patient¿s skull could support.
|