MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-554WWS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-554WWS |
Device Problem
Malposition of Device (2616)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported via phone call that the customer experienced low blood glucose of 2 mmol/l and 7.4 mmol/l which was treated with carb intake.It was also reported that the drive support cap was flushed out.It was also reported that the customer experienced high blood glucose.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
Insulin pump passed displacement test, rewind test, basic occlusion test, prime test and occlusion test.No motor error alarm noted.Driver support cap look normal.(b)(4).
|
|