BOSTON SCIENTIFIC CORPORATION INGEVITY MRI; IMPLANTABLE LEAD Back to Search Results
Model Number 7740
Device Problems Device Dislodged or Dislocated (2923); Device Contamination with Chemical or Other Material (2944); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2018
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
It was reported that this right atrial (ra) lead along with the right ventricular (rv) lead dislodged.A revision procedure was performed.During an attempt to reposition the ra lead, difficulty was encountered with retracting the helix.The ra lead was explanted and replaced.Visual inspection after explant noted tissue in the helix mechanism.The rv lead was successfully repositioned.The physician suspected potential arm movement may have been a factor in the dislodgement.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY MRI
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
GUIDANT PUERTO RICO, B.V.
no. 12, road 698
,
dorado PR 00646 3311
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8084483
MDR Text Key127569060
Report Number2124215-2018-60627
Device Sequence Number1
Product Code NVN
UDI-Device Identifier00802526523427
UDI-Public00802526523427
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/27/2020
Device Model Number7740
Device Catalogue Number7740
Device Lot Number709405
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age85 YR
-
-