• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B5LT
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Only event year known: 2018.Batch # p9419f.Device analysis: the analysis results found that b5lt device was returned with the obturator broken and outside its package unopened.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; it was noted to have a hole in the tyvek.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Additional information was requested, and the following was obtained: how did the device malfunction during the procedure? no more information is available.
 
Event Description
Malfunction.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8084730
MDR Text Key128410717
Report Number3005075853-2018-14440
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001225
UDI-Public10705036001225
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Catalogue NumberB5LT
Device Lot NumberP4T66K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Date Manufacturer Received11/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2017
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-