MAUDE Adverse Event Report: BECTON, DICKINSON AND CO. BD POSIFLUSH¿ NS FILLED SYRINGE; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND CO. BD POSIFLUSH¿ NS FILLED SYRINGE; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 306546
Device Problems Activation, Positioning or Separation Problem (2906); Improper Flow or Infusion (2954)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd posiflush¿ ns filled syringes were "only allowing about 5ml of the 10ml to be dispensed" due to difficult plunger movement.There was no report of exposure, injury, or medical intervention.
 
Manufacturer Narrative
Thirteen sample units belonging to lot number 7110587 were received for evaluation by our quality engineer team.Upon visual inspection of the samples, plunger resistance was identified.A device history record review for lot number 7110587 revealed one non-conformance during the production process that may have contributed to this incident.There was a recorded intermittent issue with the hosing in the manufacturing machinery at the time of this lot's production.Product associated with this defect was held for inspection and all affected material should have been rejected to scrap.It is possible that the product reported within this incident went undetected.A correction has been made to the silicone supply hosing machinery within the designated fill room to prevent this issue from reoccurring.Complaints received for this device and defect will be monitored by our quality team for emerging trends.
 
Event Description
It was reported that the bd posiflush ns filled syringes were "only allowing about 5ml of the 10ml to be dispensed" due to difficult plunger movement.There was no report of exposure, injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NS FILLED SYRINGE
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
MDR Report Key8086106
MDR Text Key128399086
Report Number9616657-2018-00053
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
PMA/PMN Number
K141311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Catalogue Number306546
Device Lot Number7110587
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Date Manufacturer Received10/31/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-