The subject device was returned to olympus medical systems corp (omsc).Omsc evaluated it on nov.15, 2018, and found as follows; the device was cut off at the insertion part.There were no irregularities except the cutting off of the insertion part.The manufacturing record was reviewed and found no irregularities.This type of the event is most likely related to the operator's handling.The intended use of the device is "to cut tissue within the digestive tract", not to cut the thread.Omsc assumed that the event occurred due to the user tried to cut the thread.The instruction manual of the device has already warned as follows; do not use the instrument to cut any objects other than the target tissue (e.G., stents, metallic stent wires, sewing threads, and loops).If anything other than the tissue is cut, the blades may be damaged or the cut object may be caught in the tip of the instrument, and it may become difficult to safely remove the instrument from the body.
|