Model Number 2090 |
Device Problem
Human-Device Interface Problem (2949)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer screen did not respond to the stylus.Calibrating the stylus did not resolve the issue.The programmer was returned for service.There was no patient involvement.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Manufacturer's analysis confirmed the customer comment that the programmer stylus did not align with the cursor on the screen.It was indicated that the connection between the overlay and the printed circuit board (pcb) required reseating.The programmer was repaired and returned to use.Concomitant medical products: product id 2067, serial#(b)(4).This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
|
|