It was reported that a patient underwent a laparoscopic gastric sleeve bypass procedure with a endopath xcel bladeless optiview technology w/stability sleeve and foreign material was present on the obturator tip.The device was examined under magnification for cleanliness, cracks, discoloration, nicks, etc.There was no observed foreign matter either attached or loose to the exterior or interior of the obturator's tip or any other part of the obturator.There are a couple minor scuffs/scratches near the tip that are well within the 2mm acceptability limit, that could possibly be interpreted as debris.But as this is a reprocessed device, such minor surface abrasions are common and the visual criteria accounts for these conditions.Device history record review was completed and no discrepancies were noted.There is no confirmation of any foreign matter on or in the device.Manufacturer's ref.No: pc-(b)(4).
|