Model Number 5392
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 03/05/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was able to confirm the stated reason for return of not completing its self-test and generating an error and the unit failed incoming testing (over voltage measured current and super cap voltage test, measured super cap voltage).It was additionally noted that an internal patient cable broke during the opening of the upper case (the black cable was out of the white connector).All defective parts were replaced and all other identified issues were resolved.The main board was repositioned and calibrated.The device then passed all tests with no visual or electrical anomalies.The device conformed to specifications.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) was displaying a self-test error.The epg is expected to be returned for service.There was no patient involvement.It was further reported that the epg was returned for service and subsequently tested out of specification during manufacturer's analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8102653
MDR Text Key128316520
Report Number3004593495-2018-01123
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Date Manufacturer Received11/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-