DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.705_XX
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported that during a training course it was observed that the battery reamer/drill device was getting very hot and stopped running.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.Service review: a review of the service history record indicates that the device has not been serviced for a service condition that is relevant to the current reported condition.The actual device was returned for evaluation.The battery reamer/drill device was evaluated and the reported condition that the device was getting hot and stopped running was confirmed.An assessment was performed and it was observed that the device was getting hot and making an unusual noise, the markings were faded.The reported condition that the device intermittent/stops working was not confirmed.During repair it was observed that the electronic control unit (ecu) and motor were worn, and there was an unknown debris inside the gear components.The assignable root cause was determined to be due to wear from normal use over time.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8104690
MDR Text Key128389825
Report Number8030965-2018-58466
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819491830
UDI-Public(01)7611819491830(11)131104
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705_XX
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2018
Date Manufacturer Received01/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-