Qn# (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.(b)(4) facility as part of a (b)(4) pc.Lot in july of 2015.The returned instrument was evaluated and found that the tube assembly is bent/damaged at the jaw end and the drive rod is bent/damaged where it engages the jaws and the jaws are loose and misaligned.As received , this instrument is complete and is not missing any of its components.This instrument is unusable in its current condition therefore we are able to validate this complaint.All instruments are 100% visually inspected and function tested prior to shipment to customer as this is a standardized procedure for this product line at this facility.We are unable to determine what caused the tube assembly to be damaged at the jaw end and for the tube assembly and the drive rod to be bent/damaged but mishandling of this device at the end users facility is suspected.
|