• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance which is user error/misuse/abuse.During review of the device manufacturing records for this device it was found that there were no anomalies that occurred during the manufacturing or processing of the device that would have been expected to cause or contribute to the reported event.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported that during an unspecified surgical procedure, the blade device would not turn.During an in-house engineering evaluation it was observed that the battery oscillator device blade would not turn and the device stopped running.It was further determined that the device failed pre-test for saw head housing and needle bearing corroded, check saw blade coupling, the oscillation head and set-screw were corroded and the motor got hot.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event is unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6103142063
MDR Report Key8120507
MDR Text Key128918830
Report Number8030965-2018-58590
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819491847
UDI-Public(01)7611819491847(11)151007
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Date Manufacturer Received11/13/2018
Date Device Manufactured10/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-