• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE TECHNOLOGIES CORPORATION ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE TECHNOLOGIES CORPORATION ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY Back to Search Results
Catalog Number A35864
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Patient Involvement (2645)
Event Date 10/29/2018
Event Type  malfunction  
Manufacturer Narrative
The ion torrent¿ oncomine lung cell-free total nucleic acid (cftna) research assay is labeled "for research use only.Not for use in diagnostic procedures." (ruo) no patient impact at the biocept site was reported, however, the manufacturer, thermo fisher scientific, is submitting this mdr due to a possible false positive result` for the t790m variant at the (b)(6) if the lab utilizes the ruo product to assess patient samples.The unit in question was not returned by the customer.However, a different unit from the same lot was internally investigated by the manufacturer.This investigation is ongoing.Thermo fisher scientific has initiated the recall process to remove units of ion torrent¿ oncomine lung cell-free total nucleic acid (cftna) research assay lot #: 2025404 from the market.Affected customers have been contacted to stop use of the product.
 
Event Description
A representative from (b)(6), met with a customer representative from (b)(4) to discuss technical issues regarding validation activities at the (b)(6).(b)(6) complained that a possible contaminant in the ion torrent¿ oncomine lung cell-free total nucleic acid (cftna) research assay caused the genetic variant, t790m, to show positive in sample known to be negative for the variant.This was haulting the site's validation activities.No patient samples were being analyzed when the contamination was reported.The product was not used to determine treatment or diagnostic decisions.Ion torrent¿ oncomine lung cell-free total nucleic acid (cftna) research assay is labeled "for research use only.Not for use in diagnostic procedures." the product is not a cleared or approved medical device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ION TORRENT¿ ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY
Type of Device
ONCOMINE LUNG CELL-FREE TOTAL NUCLEIC ACID (CFTNA) RESEARCH ASSAY
Manufacturer (Section D)
LIFE TECHNOLOGIES CORPORATION
7335 executive way
frederick MD 21704
Manufacturer (Section G)
LIFE TECHNOLOGIES CORPORATION
7335 executive way
frederick MD 21704
Manufacturer Contact
kelli tanzella
7335 executive way
frederick, MD 21704
MDR Report Key8122601
MDR Text Key129753557
Report Number3003335080-2018-00002
Device Sequence Number1
Product Code OTV
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/30/2019
Device Catalogue NumberA35864
Device Lot Number2025404
Was Device Available for Evaluation? No
Date Manufacturer Received10/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/30/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-