MAUDE Adverse Event Report: BIOMET UK LTD. UNKNOWN IBP HUMERAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. UNKNOWN IBP HUMERAL COMPONENT Back to Search Results
Device Problem Naturally Worn (2988)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical product - unknown ulnar stem.Therapy date - remains implanted.Report source, foreign - event occurred in (b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2018-01235.
 
Event Description
It was reported that the patient had an initial elbow arthroplasty and subsequently, the (b)(6) is requesting a custom device for a revision on an unknown date due to wear of the pe bearing.
 
Manufacturer Narrative
Cmp-0431132 this follow-up report is being submitted to relay additional information.Reported event was confirmed by review of x-rays provided.Dhr review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined, however, the information available, supported by the opinion of the operating surgeon, suggests that implant revision was required as a consequence of normal polyethylene wear after 17 years of use.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient had an initial elbow arthroplasty and subsequently the pmi group is requesting a custom device for a revision on an unknown date due to wear of the pe bearing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN IBP HUMERAL COMPONENT
Type of Device
UNKNOWN
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key8125677
MDR Text Key129079726
Report Number3002806535-2018-01236
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10
-
-