DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Device Problems Overheating of Device (1437); Intermittent Energy Output (4025)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the battery oscillator device would heat up and turn off and then work again.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device control unit did not function, the motor seized, jammed and moved heavily moving.The device also failed pretest for functional test and trigger test.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.A review of the service history record indicates that the device had been returned previously for an unrelated malfunction.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8134270
MDR Text Key129482369
Report Number8030965-2018-58749
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819491847
UDI-Public(01)7611819491847(11)141003
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Date Manufacturer Received01/24/2019
Patient Sequence Number1
-
-