MAUDE Adverse Event Report: STRYKER SPINE-US XIA TITANIUM 4.5 VITALIUM ROD 4.5 X 600 MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA TITANIUM 4.5 VITALIUM ROD 4.5 X 600 MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 48132601
Device Problems Break (1069); Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problems Dyspnea (1816); Fall (1848); Pain (1994); Injury (2348); Ambulation Difficulties (2544)
Event Date 05/16/2018
Event Type  Injury  
Event Description
It was reported that the rod broke in half and screws came out post-operatively.Patient reportedly had fall in (b)(6) of 2017 in which she tripped over a chair and fell on their knees.The patient was in pain, couldn't walk, trouble breathing and couldn't lift arm.The patient underwent revision surgery on (b)(6) 2018 to replace hardware.
 
Manufacturer Narrative
Method: surgical notes review, x-ray image review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed upon review of surgical notes and x-ray image.Products were not returned for evaluation and no lot# was provided therefore manufacturing records review and complaint history review could not be performed.Patient was reported to have had a fall in thanksgiving of 2017 in which she tripped over a chair and fell on knees.Conclusion: the root cause of the reported event could not be determined conclusively but it is likely that the patient fall (a few months ago) may have contributed to the event - perhaps a crack in the rod at the time of fall which may have lead to fracture due to fatigue over time.
 
Event Description
It was reported that the rod broke in half and screws came out post-operatively.Patient reportedly had fall in thanksgiving of 2017 in which she tripped over a chair and fell on their knees.The patient was in pain, couldn't walk, trouble breathing and couldn't lift arm.The patient underwent revision surgery on (b)(6) 2018 to replace hardware.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA TITANIUM 4.5 VITALIUM ROD 4.5 X 600 MM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8139735
MDR Text Key129574374
Report Number0009617544-2018-00290
Device Sequence Number1
Product Code NKB
UDI-Device Identifier04546540539663
UDI-Public04546540539663
Combination Product (y/n)N
PMA/PMN Number
K121342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number48132601
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received01/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age65 YR
Patient Weight175
-
-