The following was reported to gore: patient presented for treatment in the superficial femoral artery due to stenotic lesion.Atherectomy was performed, and a gore® viabahn® endoprosthesis was advanced and deployed at the intended treatment site.As reported, the distal tip broke off in the patient at some point.Another device (unknown manufacturer) was deployed to pin the distal tip against the vessel wall.More information was requested but was not made available.
|
Code 706: the engineering evaluation state the endoprosthesis, deployment knob, deployment line, nor distal tip were returned to gore.The distal shaft upon which the endoprosthesis was mounted, measured approximately 25.7 cm from the transition to tip end.The expected distal shaft length is 24.6 to 27.6.The tip end of distal shaft does not appear to have melting or interruption of the pebax material.Based on the device examination performed, anomalies were identified that could potentially be attributed to manufacturing.However, there were no additional non-conforming units beyond the one investigated in this complaint.
|