MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF THREE HOLE SHELL 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF THREE HOLE SHELL 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71336444
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/29/2018
Event Type  malfunction  
Event Description
It was reported that a patient underwent a total hip replacement.During surgery a delay greater than 30 minutes was reported due to the cup didn't fit with the liner, a bigger cup and liner size had to be implanted and fixed with bone cement.Outcome of the patient was good.
 
Manufacturer Narrative
The associated reflection liner and reflection 3 hole shell were returned and evaluated.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A quality hold was requested.With the information currently available, we are considering this to be an isolated event.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF THREE HOLE SHELL 44MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8141939
MDR Text Key129608842
Report Number1020279-2018-02713
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010460523
UDI-Public03596010460523
Combination Product (y/n)N
PMA/PMN Number
K960094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71336444
Device Lot Number15MM09496
Date Manufacturer Received11/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-