SMITH & NEPHEW, INC. REF THREE HOLE SHELL 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Catalog Number 71336444 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/29/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a total hip replacement.During surgery a delay greater than 30 minutes was reported due to the cup didn't fit with the liner, a bigger cup and liner size had to be implanted and fixed with bone cement.Outcome of the patient was good.
|
|
Manufacturer Narrative
|
The associated reflection liner and reflection 3 hole shell were returned and evaluated.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A quality hold was requested.With the information currently available, we are considering this to be an isolated event.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|