Model Number N/A
Device Problem Fitting Problem (2183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a hip arthroplasty, the liner would not seat with the acetabular shell.No adverse events have been reported as a result of the malfunction.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned device identified marks consistent with damage on the underneath of the liner.The outer radius had been indented in separate places.These indentations were consistent with the screw head used in g7 implantation.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the liner being unable to seat was attributed to the presence of screws impinging with the liner.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8148250
MDR Text Key129967185
Report Number0001825034-2018-11161
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/29/2021
Device Model NumberN/A
Device Catalogue Number010000924
Device Lot Number3929908
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2019
Was the Report Sent to FDA? No
Date Manufacturer Received08/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-