Model Number N/A |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip arthroplasty, the liner would not seat with the acetabular shell.No adverse events have been reported as a result of the malfunction.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned device identified marks consistent with damage on the underneath of the liner.The outer radius had been indented in separate places.These indentations were consistent with the screw head used in g7 implantation.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the liner being unable to seat was attributed to the presence of screws impinging with the liner.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|