INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report number: 3004608878-2018-00281.
 
Event Description
This is 1 of 2 reports.A distributor customer reported on behalf of the customer that the torque screw of the a1059 mayfield modified skull clamp came out before surgery on (b)(6) 2018.The skull clamp was in contact with patient but there was no alleged injury.The patient was prepped for surgery and a surgery delay of 30 minutes was noted.No medical revision/intervention was required.Additional information has been requested.
 
Manufacturer Narrative
The device was returned for evaluation.The heli coil was out of the thread , therefore the index knob was tight to move to lock or unlock.The device history record (dhr) shows no abnormalities relate to the reported incident found nor were there any variances, mrr¿s or reworks associated with this lot/work order number.The complaint reported was confirmed.Root cause is unknown, however the most likely reason, is parent thread had an raised indication formed by an electrical arc / rack mark, most likely caused by an insecure electrical connection between rack and power supply used by anodize supplier.The suspected small pin hole sized indication most likely prevented the helicoil from fully seating allowing the helicoil to work loose.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
kimberly shelly
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key8151470
MDR Text Key130659627
Report Number3004608878-2018-00280
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received12/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-