• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40Q
Device Problems No Device Output (1435); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Bradycardia (1751)
Event Date 12/06/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the emergency room after multiple device shocks.Upon interrogation and x-ray, it was noted that the patient's right ventricular lead had pulled back into the right atrium.The lead delivered anti tachycardia pacing followed by two shocks when it interpreted the atrial fibrillation as ventricular fibrillation.As the patient was device dependent, they suffered severe bradycardia as the implantable cardioverter defibrillator was oversensing and underpacing.The device was reprogrammed.The patient is stable with plans to revise the lead.
 
Event Description
New information received notes that the issue was resolved.The patient's right ventricular lead was explanted and a new lead was implanted.The patient tolerated the procedure well and there were no complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8154011
MDR Text Key130018386
Report Number2017865-2018-18687
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508216
UDI-Public05414734508216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberCD3357-40Q
Device Catalogue NumberCD3357-40Q
Device Lot NumberA000052762
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7122Q/58, (B)(4)
Patient Outcome(s) Required Intervention;
-
-