• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 03P75-06
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).On (b)(6) 2018 the customer reported that downloader recharger involved also became hot to the touch.Refer to apoc incident # (b)(4).Product: 04p73-04r, description: i-stat 1 d/r drc-300, kit, sn: (b)(4), mfg date: 04/06/2018.Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2018, abbott point of care (apoc) was contacted by a customer who reported that i-stat1 sn (b)(4) is displaying code 86 and is hot to the touch.The customer reports that there was no impact or delay to patient care.There were no injuries reported.The analyzer was replaced at no charge and returning for investigation.On (b)(6) 2018 the customer reported that downloader recharger sn (b)(4) also became hot to the touch that was being used during the event.There were no injuries reported on the second call.(b)(4) was replaced at no charge and also returning for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries were being used at the time of the event.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 02/12/2019.The customer reported that downloader recharger (drc) s/n (b)(4) was hot to touch when an i-stat analyzer was docked on it.The customer also reported that the cables in use were provided by apoc.The cause was determined to be defective components u12, r54, d14, q1 and q2 on the main board of the drc.A deficiency has been identified.A rocketware search spanning six months revealed no similar incidents and no evidence of a trend.No corrective or preventive action is required at this time.Analyzer s/n (b)(4) was reported to be functioning as expected by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Type of Device
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8156728
MDR Text Key131410956
Report Number2245578-2018-00314
Device Sequence Number1
Product Code CGA
UDI-Device Identifier00054749002528
UDI-Public00054749002528
Combination Product (y/n)N
PMA/PMN Number
K103195
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P75-06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2019
Date Manufacturer Received02/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-