• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CANON MEDICAL SYSTEMS CORPORATION CANON; CATHETERIZATION TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CANON MEDICAL SYSTEMS CORPORATION CANON; CATHETERIZATION TABLE Back to Search Results
Model Number CAT-870B
Device Problems Unintended Collision (1429); Unintended System Motion (1430); Unintended Movement (3026)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2018
Event Type  malfunction  
Manufacturer Narrative
Table tilted uncommanded into c- arm of system causing damage to the internal workings of the table.The patient was on the table when this occurred, but was unhurt.It was reported by the customer that the table began tilting with the patient on the table without operation intervention.They were able to get the patient off the table without injury.However, the table continued to tilt into the c-arm causing irreparable damage to the table.A replacement table has been ordered.Please note that prior to the table tilting, a nurse pierced a heparin bag that was to be hung on the iv pole on the left side of the table.The bag ruptured saturating the tilt console and the foot of the table top with the heparin.The areas were cleaned, but soon after the event (table tilt) occurred.The table, table control powerboard and connecting cable are being returned to the manufacturer for investigation.
 
Event Description
Table tilted uncommanded into c-arm of system causing damage to the internal workings of the table.Patient was on the table, but was unhurt.
 
Manufacturer Narrative
Log data was reviewed and the results of the investigation is two-fold: (a) problem with the cable connection and connection diagrams for the interlock function, which should have stopped the table from tilting and (b) table tilting was caused by a liquid spilling onto the tilt console.It was determined from the investigation by the manufacturer that liquid spilling onto the tilt console triggered a short circuit which caused the table to tilt unexpectedly.This is being considered user error because per the caution section of the operation manual the tableside console and x-ray detectors of the support units are to be covered with sterile caps to prevent contact with liquids.This was not the case.Additionally, the table tilt movement should have been stopped by the interlock function.However the interlock function was not functioning.It was found that there was a problem with the cable connection because the connection diagram was incorrect.The cable connection and the connection diagram will be modified to correct the problem on all applicable systems.The cat-870b table has been replaced at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANON
Type of Device
CATHETERIZATION TABLE
Manufacturer (Section D)
CANON MEDICAL SYSTEMS CORPORATION
1385 shimoishigami
otawara-shi, tochigi 324-8 550
JA  324-8550
MDR Report Key8170451
MDR Text Key131381504
Report Number2020563-2018-00003
Device Sequence Number1
Product Code IXR
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCAT-870B
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-