• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/02/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was not able to confirm the reported issue.Therefore the device was requested back to livanova (b)(4) for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a occlusion mechanism of a s5 roller pump was slipping during procedure.There was no report of patient injury.
 
Manufacturer Narrative
The reported s5 roller pump was returned to livanova deutschland for further investigation.During the evaluation the reported issue could not be confirmed.The device worked according the specification.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.As the issue could not be reproduced or confirmed, a root cause was not identified.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8178411
MDR Text Key131341958
Report Number9611109-2018-01415
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-