• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC Back to Search Results
Catalog Number 26-1221
Device Problem Failure to Auto Stop (2938)
Patient Problem Tissue Damage (2104)
Event Date 11/27/2018
Event Type  Injury  
Manufacturer Narrative
Udi - (b)(4).Multiple attempts to obtain additional information were not successful.Complaint sample was not returned to codman; therefore, an evaluation of the device could not be performed.A review of manufacturing records found no discrepancies when the device was released the cause(s) of the difficulty reported by the customer could not be determined.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
 
Event Description
As reported by stryker, a codman perforator failed to disengage while being used with a stryker drill, tearing the surface of the brain.The patient was able to move limbs after being wakened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
Manufacturer Contact
karen anigbo
11 cabot boulevard
mansfield, MA 02048
7819715608
MDR Report Key8178585
MDR Text Key130787653
Report Number1226348-2018-10893
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number26-1221
Device Lot NumberJ0868M
Was Device Available for Evaluation? No
Date Manufacturer Received11/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-