A tecan field service representative (fse) was dispatched to evaluate the report of the front panel falling.The fse replaced the gas spring as it was past the expected lifetime (i.E.3 years).Additionally, the fse installed a second gas spring on the front panel.Based on the upgrade configuration of the evo 100, a front panel with an adjustable shield requires 2 gas springs as specified in the service manual.At the time the instrument was upgraded by field service, only one gas spring was installed in error.A capa has been opened to investigate the issue.The evo 100 front panels are made of plexiglass and are light (3kg) and have been assessed in the past to not cause a serious injury.However, in an abundance of caution, due to prior events on larger and heavier, though similar models that resulted in moderate injuries (and for which mdrs were filed), a mdr is being filed.
|
The laboratory called to report that the door hinges collapsed and the front panel hit the operator on the head.There were no injuries (e.G.Bumps, bruises, cuts, etc) and no medical attention required.There was only one hinge from installation.
|