• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number M00542251
Device Problems Failure to Fire (2610); Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during a band ligation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the bands were twisted and would not deploy.There was no difficulty in setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during a band ligation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the bands were twisted and would not deploy.There was no difficulty in setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Problem code (b)(4) for the reportable issue of bands failed to deploy.Investigation results: only the ligator head was returned for analysis and the other components were not returned.A visual examination of the component found five bands present with some bands were moved out of their positions.It was also noticed that the ligator teeth were bent.The suture was broken, likely to separate the device from the scope.Based on the evaluation of the returned ligator head, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.It is most likely that the ligator head teeth were damaged due to handling and manipulation of the device during the procedure.Once the ligator head teeth are damaged, the suture can be detached from its position on the ligator head and this condition could have impacted the bands deployment activity which could have contributed with the reported issues.Also, it is likely that the suture was broken to separate the device from the scope and this condition is not considered as an issue of the device.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8181305
MDR Text Key130918643
Report Number3005099803-2018-61837
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/13/2019
Device Model NumberM00542251
Device Catalogue Number4225
Device Lot Number0022371482
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Date Manufacturer Received01/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age35 YR
Patient Weight80
-
-