The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.Reporter of the event indicated that the device serial number was not known.Device labeling addresses the reported event as follows: precautions: antiemetics, antispasmodic, and anticholinergic drugs may be prescribed to lessen the early placement symptoms such as nausea, vomiting, and abdominal pain.Patients will need to immediately contact their physician for any severe or unusual symptoms.Placement of the balloon within the stomach produces an expected and predictable reaction characterized most commonly by a feeling of heaviness in the abdomen, nausea and vomiting, gastroesophageal reflux, belching, esophagitis, heartburn, diarrhea and, at times, abdominal, back or epigastric pain and cramping.Food digestion may be slowed during this adjustment period.These symptoms can be treated with antiemetic, antispasmodic, and anticholinergic medications.Typically the stomach acclimates to the presence of the device within the first 2 weeks.In order to prevent or ameliorate the symptoms most frequently experienced during the adjustment period, it is recommended that the physician use proton pump inhibitors (ppis), antiemetics, antispasmodics, and anticholinergic medications prophylactically (before orbera¿ placement).Patients should be advised to immediately contact their physician for any unusually severe or worsening symptoms.The physiological response of the patient to the presence of orbera¿ may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response.Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients need to be evaluated and the device removed at or within 6 months of placement.Clinical data does not exist to support use of an individual orbera¿ beyond 6 months.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications that may result from the use of this product include the risks associated with the medications and methods utilized in the endoscopic procedure, the risks associated with any endoscopic procedure, the risks associated with the orbera intragastric balloon specifically, and the risks associated with the patient's degree of intolerance to a foreign object placed in the stomach.Possible complications - possible complications of the use of orbera¿ include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach, delayed gastric emptying and/or the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus.Abdominal or back pain; either steady or cyclic.Spontaneous over inflation.Warnings: spontaneous over inflation of an indwelling balloon has been reported in patients with orbera¿.Symptoms of balloon over-inflation include intense abdominal pain, swelling of the abdomen (abdominal distension) with or without discomfort, difficulty breathing, and/or vomiting.Patients experiencing any of these symptoms should be counseled to seek immediate care.
|
Supplement #1 - medwatch sent to the fda on 25/feb/2019.Additional information: device evaluation summary: the device was returned to apollo, and a visual inspection was performed.A portion of the balloon shell was noted to be discolored, as approximately 20% of the shell was light brown in appearance.Yellow and white particles were observed on the outer surface of the shell and center patch.A valve test was performed, and when di water was injected into the balloon valve, the flow of fluid was continuous and unobstructed.An air leak test was performed, and leakage was noted from several openings on the radius of the balloon shell.Under microscopic analysis, five openings were noted on the radius of the shell.All five openings on the balloon shell were noted to have striated edges, consistent with damage from a surgical tool.Orange and black particles were noted on the inner surface of the valve channel.
|