• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot g346 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot g346 for the reported issue shows no trends.Trends were reviewed for complaint categories, tubing leak and alarm #18: system pressure.No trends were detected for each complaint category.Photographs were returned for evaluation.A review of the customer provided photographs confirm the tubing leak as blood is seen in the pump head location and pump tubing loop.The damage to the tubing is consistent with damage caused by improper installation of the pump loop tubing into the pump head.During installation of the kit by the end user if the pump loop is forcefully rubbed against the pump head it can result in scraping and potentially cutting the tubing loop as the pump head rotates.A material trace of the tubing used to manufacture lot g346 found no related non-conformances.A device history record review did not identify any related non-conformances, and this kit lot had passed all lot release testing.Cellex kits are 100% leak tested prior to packaging.A leak of this nature would have been detected during in-process testing; therefore, it is unlikely the tubing damage was present at the time of manufacture.The investigation determined the root cause of the tubing leak was most likely improper installation of the pump loop by the end user.No manufacturing related defects were identified through this investigation.No further action is required at this time.This investigation is now complete.(b)(4).
 
Event Description
The customer called to report a tubing leak during the treatment procedure.The customer stated they received an alarm #18: system pressure alarm at 321 ml whole blood processed and noticed a leak coming from the collect pump tubing segment.The customer stated there appears to be a slit in the tubing segment.The customer aborted the procedure and did not return blood to the patient.The customer stated the patient was stable.The customer has returned photographs for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key8182671
MDR Text Key131796697
Report Number2523595-2018-00174
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)G346(17)200701
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2020
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberG346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient Weight93
-
-