MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Model Number LSMU1350726
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Patient Involvement (2645)
Event Date 11/26/2018
Event Type  malfunction  
Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.(expiry date 06/2020).
 
Event Description
It was reported that during treatment for a perforated left common iliac artery via access through the right common femoral artery, the balloon expandable vascular covered stent delivery system allegedly failed to insert through the valve of the 6fr introducer sheath.Another device was used to complete the procedure.There was no reported patient contact.
 
Event Description
It was reported that during treatment for a perforated left common iliac artery via access through the right common femoral artery, the balloon expandable vascular covered stent delivery system allegedly failed to insert through the valve of the 6fr introducer sheath.Another device was used to complete the procedure.There was no reported patient contact.
 
Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Investigation summary: the investigation is unconfirmed for the device incompatibility failure mode reported.While resistance was encountered, the device was successfully inserted into the sheath valve and passed through a 6fr terumo radifocus introducer ii sheath.It is unknown if there were procedural or handling techniques that contributed to the reported event.The manufacturing documentation was reviewed and shows no anomalies with the manufacture of this lot to suggest a product issue.Labeling review: the ifu for the lifestream product was reviewed and contains the following information relevant to the reported event: precautions: the device should only be used by physicians who are trained in endovascular procedures and are familiar with the complications, side effects and hazards of peripheral vascular interventions.Prior to device use, refer to the covered stent sizing table on the label and read the instructions for use.Crossing the implant with catheters or other adjunct devices can result in covered stent dislodgement or damage.Potential patient/device adverse effects that may occur include, but are not limited to, the following: covered stent dislodgement from balloon during tracking procedure.Covered stent misplacement during placement procedure.Directions for use: site access and preparation: using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.Covered stent size selection select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation carefully remove the selected device from the package.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system air evacuation a 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent advance the endovascular system over the guidewire into the introducer sheath.(expiry date 06/2020), date rec¿d by mfr.Device evaluated by mfr (results 1).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT
Type of Device
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
MDR Report Key8182791
MDR Text Key131377074
Report Number9616666-2018-00180
Device Sequence Number1
Product Code PRL
UDI-Device Identifier05391522081130
UDI-Public(01)05391522081130
Combination Product (y/n)N
PMA/PMN Number
P160024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLSMU1350726
Device Catalogue NumberLSMU1350726
Device Lot NumberCMBU0314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Date Manufacturer Received01/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age68 YR
Patient Weight58
-
-