Model Number 802018 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) was unable to duplicate the reported complaint.Since the issue was intermittent, as a precaution, he replaced the flow sensor and flow module.The suspect parts will be returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there were multiple occurrences of the flow sensor intermittently not displaying a flow reading.The sensor was unplugged then plugged back in and it worked correctly.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per supplier evaluation, the board was tested on a flow loop for approximately 18 hours.The problem of not reporting flow after ten minutes was not observed.The flow board was cycled ten times and was reporting flow.After waiting 15 minutes it never failed to report flow.The problem could not be reproduced.Further laboratory evaluation was initiated and the product surveillance technician (pst) observed the flow module and sensor to operate for four days with no loss of flow display.A pin was found to be bent on the connector and was straightened before re-testing.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the flow sensor to operate as intended throughout the evaluation.He did find that the application board in the flow module was defective and was the cause of the intermittent flow readings.
|
|
Search Alerts/Recalls
|