Model Number 595000-001
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining function.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that a patient on the freedom driver at home felt lightheaded and called out for help.The customer also reported that the patient's freedom driver stopped without any alarms then restarted with alarms.The customer also reported that the patient's caregiver quickly switched the patient to a backup freedom driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
The customer-reported issue of the driver pausing/stopping and restarting with a fault alarm was confirmed via review of the driver's alarm history and testing.The alarm history was reviewed and revealed alarm codes produced as a result of the primary motor not engaging for greater than or equal to 5 seconds after continuous open or short detected.During investigation testing, the driver started up on the secondary motor and enunciated an alarm as designed.The root cause of the customer-reported issue was determined to be a malfunction of the main printed circuit board assembly (pcba) not powering the primary motor piston cylinder assembly (pca) triggering the driver to convert to secondary motor operation.This led to engagement of the secondary motor, which aligns with the reported brief pause in operation while switching over, and then starting back up with a fault alarm.Syncardia has a corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation and is closing this file.Ce 4592 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8186733
MDR Text Key131311580
Report Number3003761017-2018-00555
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Date Manufacturer Received11/29/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age50 YR
-
-