Model Number N/A |
Device Problems
Difficult to Insert (1316); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the liner did not seat properly into the acetabular shell.No adverse events have been reported as a result of the malfunction.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection identified damage to the locking feature, deformation on the scallops and a gouge to the scallop.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|