C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI14
Device Problems Inaccurate Flow Rate (1249); Improper Flow or Infusion (2954)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that there was no urine flow on the day of placement.
 
Event Description
It was reported that there was no urine flow on the day of placement.
 
Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.The evaluation found no blockage observed in the drainage lumen.Water was introduced through the drainage eye and exited from the drainage funnel without difficulty.The sample was sent for flow rate test and passed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿[(6)since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.(7)when urinary flow cannot be noted, confirm that the catheter is neither occluded nor broken.(8)avoid force on the connecting parts as they may be accidentally disconnected due to the weight of the drainage bag etc.And may cause urine spill.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8189718
MDR Text Key131311277
Report Number1018233-2018-06112
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016745
UDI-Public(01)00801741016745
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2021
Device Model Number0165SI14
Device Catalogue Number0165SI14
Device Lot Number8EZ0098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2018
Date Manufacturer Received01/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-