• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number 05.001.080
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2018
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the air pen drive device was heating up and locking up.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown, however it was reported that the event occurred on (b)(6) 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Service history review: the service history review shows no previous service conditions relevant to the current complaint issue.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device had seized, jammed and moved heavily.The device also failed pretest for check external leak tightness and check internal leak tightness.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR PEN DRIVE
Type of Device
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8189793
MDR Text Key131175811
Report Number8030965-2018-59194
Device Sequence Number1
Product Code HBB
UDI-Device Identifier7611819830899
UDI-Public(01)7611819830899(11)130503
Combination Product (y/n)N
PMA/PMN Number
K093361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Date Manufacturer Received01/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-