• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome membrane leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.No kit lot number was provided; therefore, no batch record review could be performed.Trends were reviewed for complaint categories, pressure dome membrane leak and kit installation/removal failure.No trends were detected for each complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Section 9-6 of the therakos¿ cellex¿ photopheresis system operator's manual states "the integrity of pressure monitoring will be compromised if the pressure domes are removed after prime." additionally section 4-18 and section 9-9 of the therakos cellex photopheresis system operator's manual states: "do not remove any pressure domes during a treatment.Please call the therakos hotline for assistance as needed during troubleshooting events." a review of the customer's complaint description identified that the customer had removed or manipulated the system pressure dome during the treatment.The root cause of the pressure dome membrane leak is most likely the end user removing the pressure domes after prime was completed.The customer was verbally informed not to remove the pressure domes once a prime is completed.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.Mc: (b)(4).(b)(6) 12/21/2018.
 
Event Description
The customer called to report a pressure dome membrane leak during the treatment procedure.The customer stated to save time they loaded a kit onto the instrument and primed the instrument outside of the treatment room.The prime was completed, then the nurse removed the pressure domes from the pressure transducers and powered off the instrument.The customer stated they then moved the instrument into the treatment room, reinstalled the pressure domes, and powered the instrument on.The customer stated within 5 minutes of starting the treatment they noticed a pressure dome membrane leak around the centrifuge pressure sensor.The customer aborted the treatment and did not return blood to the patient.The customer stated the patient was stable and returned the following day for treatment.The customer did not return product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key8189853
MDR Text Key131386048
Report Number2523595-2018-00178
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-