• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100A VENTILATOR; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100A VENTILATOR; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems Inaccurate Flow Rate (1249); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2018
Event Type  Injury  
Manufacturer Narrative
The customer reported evaluating the device and determined that the pressure regulator (item773913) flow was out specifications and was the likely cause of this event.A replacement pressure regulator 773913 was ordered.In the event that the device/component is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported a loud noise and vibration, while in patient use on this oscillator device.The customer stated that the patient was removed and placed on an alternate device with no harm or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100A VENTILATOR
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8190060
MDR Text Key131173130
Report Number2021710-2018-09702
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/16/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-