• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number SSVO
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device was a newly opened one and it gave out weird noise when attached to the patient's tracheostomy tube.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: the sample was received for evaluation and no failure was confirmed.A visual inspection was performed and no cuts, deformities or damage was observed in the membrane nor the diaphragm, and the membrane is in the right position.Received sample met with the product specifications and no corrective actions are required.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key8192008
MDR Text Key131305511
Report Number2936999-2018-00789
Device Sequence Number1
Product Code BTO
UDI-Device Identifier40884522005372
UDI-Public40884522005372
Combination Product (y/n)N
PMA/PMN Number
K945407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/25/2022
Device Model NumberSSVO
Device Catalogue NumberSSVO
Device Lot Number17A0574JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2019
Date Manufacturer Received04/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-