ROCHE DIAGNOSTICS ELECSYS FT4 II ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07027397190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
The customer complained of questionable elecsys tsh assay, elecsys ft4 ii assay, and elecsys ft3 iii results for 1 patient tested on a cobas 8000 e 801 module compared to the results from an abbott architect.This medwatch will cover ft4 ii.Refer to medwatch with patient identifier (b)(6) for information on tsh.Refer to medwatch with patient identifier (b)(6) for information on ft3 iii.The erroneous results were reported outside of the laboratory.There was no allegation of an adverse event.The cobas e801 serial number used as the customer's site is (b)(4).The serial number for the cobas e801 used at the investigation site is (b)(4).The tsh reagent lot used on this cobas e801 was 283556 with an expiration date of feb-2019.The ft4 ii reagent lot used on this cobas e801 was 303203 with an expiration date of apr-2019.The ft3 iii reagent lot used on this cobas e801 was 314824 with an expiration date of may-2019.The investigation determined that in relation to the differences of the tsh and ft3 and ft4 values generated with the analyzers from roche diagnostics and abbott: it needs to be taken into account that assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used.For the tsh and ft3 and ft4 assays, all generated values ¿ with the analyzers from roche diagnostics and abbott ¿ fit with each other, from a location of these values versus the respective normal reference ranges of the assays.From the information provided and the analysis thereof, a general reagent issue most likely can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 II ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key8195640
MDR Text Key131494452
Report Number1823260-2018-05097
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07027397190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age28 YR
-
-