The customer complained of questionable leukocytes, nitrite, protein, glucose, and urobilinogen results for 1 patient urine sample tested with chemstrip 10 md urine test strips on a chemstrip 101 urine analyzer serial number (b)(4).Of the data provided, there were discrepant nitrite results.The initial nitrite result was positive.The repeat nitrite result was negative.There was no allegation of an adverse event.The calibration and qc were acceptable.The suspect product was requested to be returned for investigation.
|
The customer material was returned for investigation.The retention material of lot 32223100 and the customer material of lot 32223101 were both checked on a cobas urisys 1800 / cobas u411 with native urine, a nitrite dilution series, a protein dilution series, a glucose dilution series, and a leucocytes dilution series.The investigation did not identify a product problem.The cause of the event could not be determined.
|