An event of embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined; however, information from the field indicated that the user believed this was most likely due to mis-sizing.
|
On (b)(6) 2018, a 30mm amplatzer atrial septal occluder (aso) was selected.During the procedure, the 30mm aso deployed in a cobra shape and per the user, the deformation event did not have the potential to cause patient harm or injury.The device was exchanged for a second 30mm aso, which was considered a mis-sizing so the device was exchanged for a 32mm aso.The 32mm aso was deployed and released, but embolized after implantation.Again, per user, the embolization event was most likely due to mis-sizing of the device.The patient was referred to cardiac surgery where the embolized device was retrieved and the atrial septal defect was closed with a patch.The patient remained hemodynamically stable throughout the procedure and is reported stable.
|